1 I-MP1900A I-Bert Stress Calibration Imikhuba Engcono Kakhulu

I-MP1900A I-Bert Stress Calibration Imikhuba Engcono Kakhulu

PCI Express 6.0 CEM:
BERT Stress Calibration
Izindlela Ezinhle Kakhulu

ISIFINYEZO SEZOBUCHWEPHESHE

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

Isingeniso

The PCI Express (PCIe) CEM Specification requires testing both transmitters and receivers to ensure that devices comply with the specification. Tektronix  provides automated BERT receiver (RX) stress calibration  and receiver test software for PCIe 6.0 CEM testing.  This software automates the calibration process for a  system composed of a Tektronix DPO70000SX real time  oscilloscope and Anritsu MP1900A Bit Error Rate Tester  (BERT). Features are implemented in TekRXTest Software  publicly available today supporting RX characterization,  debug, pre-compliance and compliance testing per the 6.0  CEM specification.

TekRxTest software offers a wizard-based approach  to intuitively guide users through the complex receiver  calibration process. The process, when done manually, can  easily take two to three days, while with this new solution  calibration can typically be completed in hours.

TekRxTest implemented the following unique approaches to make it more efficient than other available techniques:

  • Patented Tx EQ calibration shortens the process from several hours on comparable systems to ~30 mins, without compromising accuracy required by the specifications.
  • Hardware accelerated FastFrame™ approach shortens the waveform averaging process required in the TP2 calibration process by hours.

In addition, TekRxTest integrated receiver calibration and testing supports testing for all generations of the PCI Express standard.

Ingemuva

PCIe receiver testing specifies the calibration of test signals. During actual operation of a high-speed serial link, numerous impairments will close an eye. PCIe standards  offer a well-defined CEM Test Spec that specifies  combinations of impairments to use for testing. Stressed  eye targets are defined with the full pad-to-pad channel,  with Compliance Base Board (CBB) and Compliance Load  Board (CLB) in the path through the CEM connector. The  BERT PPG acts as a surrogate PCIE transmitter in the  scenario of receiver testing. Once the recipe for the  stressed eye target is determined, one removes the  fixturing that isn’t used in testing — CLB for Add-In-Card  testing, or CBB for System testing. This effectively refunds  the loss budget permitted for the device-under-test (DUT).

Users must perform BERT stress calibration before RX  tests are conducted on a DUT. This BERT calibration  ensures the magnitude of stresses applied to the DUT  RX are in accordance with PCIe 6.0 spec limits. This RX  testing determines RX pass/fail and margin performance.  During BERT stress calibration, the user controls multiple  parameters during the calibration run. Receiver stressed  eye calibration is a complex, iterative process and human  error is a significant risk. As such, manual calibration is  unsustainable for routine testing. This technical brief  discusses those controls and provides recommendations  to maximize the probability of successful calibration  after all steps are complete. Guidance is given on when  different modes should be used, and troubleshooting tips  are provided.

2 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

TekRXTest RX Software

TekRXTest is a software framework that supports automated BERT stress calibration for multiple technologies including PCIe 6.0 CEM. Umfanekiso 1 shows the start screen for TekRXTest, and the RX personalities supported as of May 2025.

More info on TekRXTest is available lapha.


Figure 1. Start screen for TekRXTest and RX personalities.

WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

BERT Stresses For PCI Express 6.0 CEM

BERT stress calibration for PCIe 6.0 RX testing is broken into two parts: short channel (TP3, Umfanekiso 3) and long channel (TP2,  Umfanekiso 2). TP3 signal validation is performed first with BERT output signals connected directly through RF coaxial cables to  scope inputs. BERT tests using TP3 connections are pulse width jitter, differential amplitude, AC-DC balance, preset, random  jitter (RJ) and sinusoidal jitter (SJ). 6dB external attenuators should be installed on the front of each oscilloscope ATI channel  to ensure the best combination of voltage range and noise performance. These attenuators are available as part of the  DPO7RFKx connector kits from Tektronix.

iso

Figure 2. Long channel BERT stress calibration for PCIe 6.0 RX testing.

ububanzi

Figure 3. Short channel BERT stress calibration for PCIe 6.0 RX testing.

4 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

Once TP3 calibrations are complete, a differential loss ISI channel is added between the BERT PPG and scope inputs. An  example connection diagram is shown in Umfanekiso 4. TP2 calibrations are performed including insertion loss (IL) verification,  common-mode interference (CMI) and differential mode interference (DMI), optimal preset and CTLE selection and stressed  eye. Final stressed eye parameters are compared to spec limits to determine successful completion of the BERT calibration.

isiteshi

Umfanekiso 4. Isbample connection diagram showing a differential loss ISI channel added between the BERT PPG and oscilloscope inputs.

WWW.TEK.COM | 5

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

User Controls During BERT Calibration

TekRXTest Software prompts the user to specify some settings prior to performing TP3 and TP2 BERT stress calibration.  Ithebula 1 lists these controls, their description, and how to set these values to complete a successful calibration.

Table 1. List of parameters set during calibration.

Note 1: Versions of Sigtest & SEASIM will be updated when fully tested and recommended by PCI-SIG.

Note 2: Although not required for PCIe 6.0 CEM-RX pass/fail testing, the user may enable JTOL testing with up to 14 SJ separate frequency tones to perform RX  margin testing. These SJ tones must be calibrated during TP3.

Izilungiselelo zomsebenzisi:

Ingxenye

Ipharamitha

Okuzenzakalelayo

Kunconyiwe

Amanothi

RT scope

 Positive ch

Ch1

 Ch1

RT scope

 Negative ch

Ch2

 Ch2

TekRXService

 Isikhathi siphelile

300 isekhondi

 300 isekhondi

TekRXService

Sigtest  

inguqulo

6.1.10

 6.1.10

Inothi 1

TekRXService

Seasim  

inguqulo

2.0.103 f

2.0.103f

Inothi 1

TP3 attenuator  settings

Acquisition  mode

Ngokushesha  

Uhlaka

Fast Frame

TP3 attenuator  settings

Susa ukushumeka

 Ikhutshaziwe

 Ikhutshaziwe

TP3 attenuator  settings

Ch1  

ukunciphisa

6db

 6db

TP3 attenuator  settings

Ch2  

ukunciphisa

6db

 6db

I-TP3  

ukulinganisa

Ampukufunda

800 mV

 800 mV

I-TP3  

ukulinganisa

RJ

I-0.25 ps  

I-RMS

0.25 ps RMS

I-TP3  

ukulinganisa

SJ @  

100 MHz

1.563 ps

 1.563 ps

I-TP3  

ukulinganisa

Multi-tone  Calibration

Ikhutshaziwe

Kuchazwe umsebenzisi

Inothi 3

TP2 attenuator  settings

Acquisition  mode

Ngokushesha  

Uhlaka

Fast Frame

TP2 attenuator  settings

Susa ukushumeka

 Ikhutshaziwe

 Ikhutshaziwe

TP2 attenuator  settings

De-embed  (channel)

Ikhutshaziwe

 Ikhutshaziwe

TP2 attenuator  settings

Ch1  

ukunciphisa

6db

 6db

6 | WWW.TEK.COM

Ingxenye

Ipharamitha

Okuzenzakalelayo

Kunconyiwe

Amanothi

TP2 attenuator  settings

Ch2  

ukunciphisa

6db

 6db

TP2 DMI and  

I-CMI

I-DMI

15 mV

 15 mV

TP2 DMI and  

I-CMI

I-CMI

75 mV

 75 mV

TP2 stressed  eye params

# acqs

 3

 3

TP2 stressed  eye params

Ububanzi bamehlo  

okuhlosiwe

I-3.125 ps

 I-3.125 ps

System  Host

TP2 stressed  eye params

Eye height  target

6 mV

 6 mV

System  Host

TP2 stressed  eye params

DMI min

 5 mV

 5 mV

System  Host

TP2 stressed  eye params

DMI max

 25 mV

 25 mV

System  Host

TP2 stressed  eye params

SJ min

 I-1 ps

 I-1 ps

System  Host

TP2 stressed  eye params

SJ max

 I-3 ps

 I-3 ps

System  Host

TP2 stressed  eye params

Amplitude  min

720 mV

 720 mV

System  Host

TP2 stressed  eye params

Amplitude  max

800 mV

 800 mV

System  Host

Remote access  – Config

I-IP yendawo  

ikheli

I-PC yokusingatha  

IP

Host PC IP

Remote access  – Config

Ukulalela  

itheku

4004

 4004

Remote access  – Config

Isikhathi siphelile

20 isekhondi

 20 isekhondi

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

Launch & Run TP3 Calibration

When a user starts a short channel calibration, the RX software provides a connection diagram to ensure signals are connected properly from the BERT PPG to the scope inputs. Proper care should be used in making RF connections between cables and components. (Details are not covered in this technical brief). Refer to the High Performance Oscilloscopes Best  Practices Manual futhi DPO70000SX Quick Start User Manual, and use good lab practices. For assistance making proper  connections, please contact your local Tektronix field application engineer.


Figure 5. TP3 calibration connection diagram.

WWW.TEK.COM | 7

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

With RX software launched and “connected” via IP address to scope and BERT, the user may INITIALIZE equipment for TP3  calibration. The RX software will automatically initialize instruments when the user clicks the INITIALIZE EQUIPMENT button  within TP3 calibration test flow. Initialization of the scope and BERT takes less than five minutes. By default, the scope will  perform signal validation during the initialization process. To speed up initialization, the user may select the check-box to  SKIP SIGNAL VALIDATION DURING INITIALIZATION. These controls are shown in Umfanekiso 6 ngezansi.


Figure 6. TP3 calibration connection wizard showing the Initialization steps.

The user may run all TP3 calibrations automatically, or one at a time. It is recommended that new users manually perform and  observe results for each calibration step. Manual mode is also recommended if hardware setup is new to ensure each TP3  calibration test completes correctly. For subsequent stress calibration runs, auto mode may be considered.

Baseline pulse width jitter (PWJ) is needed during TP3 to provide a reference for Seasim. During the PWJ Calibration  step, the user may opt for the scope to measure PWJ (default), or enter known values for PWJ RJ and PWJ DJ for their  specific setup.

When PWJ Calibration is complete, the next three calibration steps will proceed without user input (AC-DC balance,  Amplitude, Preset). During the Preset calibration step, all presets (Q0 to Q9) are calibrated. An IL measurement is performed  after Preset and is required only if the user plans to perform IL verification step during TP2. If the user will use VNA loss  measurements to establish and adjust ISI channel losses during TP2 calibration, TP3 IL measurement step can be skipped to  save time. If the IL measurement step is performed, the scope will measure short-channel loss (specify # scope acquisitions  and # averages) using step response from BERT and analysis via Seasim. The user may alternatively manually enter in the  short-channel loss (dB at 16 GHz) if known.

8 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices

ISIFINYEZO SEZOBUCHWEPHESHE

The next three calibration steps will proceed without user input (RJ, SJ, SJ@210MHz). After the last TP3 calibration step is  complete, the user may provide information to identify the calibration: Unique ID, Generated by, and comments field. Click  the SAVE button to save this TP3 calibration in the RX software for future use.

Multi-tone SJ Calibration

The user may enable multi-tone SJ calibration during TP3 to perform JTOL margin testing. The multi-tone SJ feature is  disabled by default but may be enabled in SETTINGS ➞ TP3 Calibrations via checkbox. Longer calibration times will be  incurred the more SJ frequencies that are specified (about five to seven minutes per SJ tone), and the higher number  calibration points tested between Min and Max SJ values. Even after passing compliance testing, it may be desirable to test  design margins. TekRXTest software makes it straightforward to retroactively add multi-tone SJ calibration to an existing  TP3 calibration.


Figure 7. Configuring multi-tone SJ frequency and ampukufunda.

The multi-tone SJ feature supports:

• SJ sweep across frequencies from 30 kHz up to 100 MHz.

• ≤14 SJ frequencies.

• Two to eight calibration points per SJ frequency.

• Min SJ value determined by BERT hardware intrinsic jitter.

• Max SJ value determined by BERT hardware SJ limits.

WWW.TEK.COM | 9

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

TP3 Calibration Time

TP3 calibration takes about 45 minutes using default software settings. This run-time assumes hardware and software prerequisites are met, proper communication exists between the host PC running TekRXTest and the test instruments, correct connections are made per connection diagrams and instruments are in good working order.

Several factors allow TP3 calibrations to run faster:

• Skip signal validation during initialization.

• Manually enter BERT PWJ values.

• Skip IL measurement if using VNA to measure ISI channel loss in TP2.

• Use auto-calibration mode.

Launch & Run TP2 Calibration

Several factors will cause longer TP3 run-time including:

• Signal validation during initialization.

• Performing BERT PWJ measurement.

• Performing short-channel IL measurement.

• Using manual run mode, and pausing between  calibration steps.

The same RF cables used to connect the BERT PPG to the scope inputs during TP3 calibration should remain in place when  performing TP2 calibration. In the shift from TP3 to TP2 setup, these RF cables are disconnected from scope inputs and  connected to PCI-SIG 6.0 CEM test fixtures. These connections are shown in Umfanekiso 8 ngezansi.

 

Figure 8. TP3 and TP2 RF cable connection.

Isixhumi WORD? I-Attenuator

1.85 mm to 2.92 mm Adapter

10 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

The user must perform and save a TP3 short channel calibration prior to running a long channel calibration. TP2 stress  calibration will prompt the user for DUT type (System Host or Add-in Card) and the TP3 calibration to use for TP2 calibration.  Similar to TP3:

1. TekRXTest software provides a connection diagram to ensure signals are connected properly from BERT PPG to the  scope inputs. See Umfanekiso 8.

2. The user will initialize instruments using INITIALIZE EQUIPMENT button before performing TP2 calibration steps.

3. TP2 calibration settings are provided in the SETTINGS menu. These user settings are shown in Table 1.

4. TekRXTest software provides BERT calibration steps that may be run manually (one at a time) or in auto mode.

5. TP2 test report with plots and BERT calibrated values are available upon completion.

ISI Channel Loss

The user should acquire approved PCIe 6.0 fixtures including ISI board and cables from PCI- SIG to enable signal path  connections shown in Umfanekiso 8. VNA insertion loss measurements are recommended on ISI channel fixtures and associated  cables to ensure the prescribed amount of loss is in place during TP2 calibration.

Target ISI channel loss depends on whether a System Host or Add-in Card is being tested. For System Host, the total ISI  channel loss target is 24 dB @ 16 GHz (Add-in Card is 28 dB @ 16 GHz). These targets are shown in the RX software at the IL for  DMI/CMI calibration step shown in Umfanekiso 9 ngezansi.

ukulinganisa

Figure 9. TP2 calibration targets.

WWW.TEK.COM | 11

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

PCIe 6.0 ISI channel loss boards provide traces with about 0.5 dB loss increments to allow for adjustment. The user will vary  channel loss together with other BERT stress parameters to meet stress calibration closure requirements. Because each ISI  board will vary in its loss profile, it is recommended the user measure ISI channel loss on the fixture to be used prior to TP2  calibration. The goal is to identify a cable + ISI channel pair that exhibit the target channel loss (24 dB @ 16 GHz for System  Host, 28 dB @ 16 GHz for add-in card). It is Tektronix’ experience that lanes meeting ISI channel requirements do vary. Actual  convergence during the March 2025 pre-FYI plugfest was obtained using ISI pair 2 and 14 for System Host, ISI pair 23 and 0 for Add-in Card. This lane variation is dependent on RF cable, connector and fixture loss differences.

TekRXTest software will embed the appropriate package loss onto the scope’s acquired waveform during TP2 calibration so that the step waveforms passed to Seasim represent the entire 30-33 dB ISI channel loss required for CEM compliance testing. This total loss is shown in the IL Measurement in Umfanekiso 10.

iqoqo

Figure 10. Total package loss.

Once an ISI channel is chosen, TP2 stress calibration can begin. The recommendation from PCI- SIG is to begin with max  ISI channel loss for initial TP2 calibration run. The user may perform the IL measurement for DMI/CMI step, but this step is  optional and may be skipped if VNA measured data was used to verify target channel loss. DMI and CMI calibrations will run  without any user input. At the IL Measurement step, the user may select MANUAL ENTRY and enter in the total ISI channel  loss (i.e. VNA measured loss + embedded package loss). For System Host this target is 30-33 dB.

12 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

Optimal preset calibration is performed next shown by Umfanekiso 11. The user may use the CONFIGURATION button to determine  what presets are considered during this step (Q0 to Q9). As shown in Umfanekiso 11, presets Q5, Q6, Q8, Q9 were selected based on  recommendation from PCIe 6.0 PHY Test spec working group. In Tektronix’ experience, successful calibrations have been  achieved using Q5, Q7 and Q8. The user may control the number of acquisitions for optimal preset step. The default number of acquisitions is three, but this value may be increased to ≥5, resulting in more waveform averages used to for preset eye area calculation.

itheku

Figure 11. Optimal preset calibration.

WWW.TEK.COM | 13

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

TekRXTest software will choose the preset resulting in the largest average eye area. In the example eboniswe ku Umfanekiso 12,  preset Q5 was found to yield the highest eye area of 5.274. In this example, the RX software will use Q5 in the final Stressed  Eye Cal step.

In some cases, more than one preset may result in a large eye area. To choose the best preset, the ratio of eye width /  eye height should be evaluated. With these ratios computed, the preset resulting in the ratio being closest to 50% is  recommended. An example inikezwe ngezansi:

Converged eye #1: Converged eye #2:

Preset = Q5 Preset = Q7

Eye width (EW) = 1.484 Eye width (EW) = 1.645

Eye height (EH) = 3.554 Eye height (EH) = 3.487

Ratio = EW / EH = 0.42 Ratio = EW / EH = 0.47

Choose this setting

Figure 12. TP2 calibration optimum CTLE and preset selection example. Note that the “Bad” results for Q8, Q9 presets are normal in this case, when searching for the  optimum CTLE.

Final calibration step is Stressed Eye Cal. This step uses the preset determined in the previous step yielding the largest eye  area. With this preset configured in the BERT, a step waveform is captured and averaged then sent to Seasim for analysis.  Tektronix has implemented HW- assisted 10Ktrace averaging across 960 UIs to accelerate this process. Combinations of SJ,  DMI and amplitude are swept in Seasim. Tie-breakers are used to make decisions in cases where different values result in  similar final stressed eye conditions.

14 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

TekRXTest software will use target values defined in TP2 STRESSED EYE PARAMETERS to adjust variable parameters to  meet eye target value. The default number of waveforms for Stressed Eye Cal is three, but this value may be adjusted by the  user. (A higher number of acquisitions yields more data but takes longer to acquire and process.) Tektronix recommends to  leave this value set to # acqs for stressed eye = 3 for initial TP2 runs. If three acquisitions are not sufficient, increase to ≥5 to  achieve stable results.

Nominal eye height and eye width targets for PCIe 6.0 are:

• Eye width: 3.125 ps (range of 2.825 ps to 3.425 ps)

• Eye height: 6 mV (range of 5.5 mV to 6.5 mV)

phini

Figure 13. Configuring number of waveforms averaging for stressed eye.

If the RX software cannot find a converging set of variable stresses after sweeping all available parameters, the user may take several actions:

• Change ISI lane pairs to reduce channel loss by 0.5 dB. This will require a re-run of TP2 calibrations with reduced  channel loss.

• In the case of changing the ISI pair, the user should rerun the two steps that are dependent on the ISI pair: the Preset  Selection, and the Stressed Eye Cal. These should be rerun to 1) find the optimal Preset and CTLE based on the current ISI  pair, and 2) find the final stress recipe based on the current ISI pair.

Consider using manual sweep settings while holding other variables constant. An addendum will be added to this document  providing guidance on how best to use sweep mode in this case.

WWW.TEK.COM | 15

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices ISIFINYEZO SEZOBUCHWEPHESHE

TP2 Calibration Time

TP2 calibration takes two and a half to four hours to run using standard software settings. This run time is affected by  several factors including:

• Is IL Verification step run, or is VNA data used?

• Is ISI channel loss within target range?

• Can convergence be obtained with combination of BERT variable stresses and ISI channel loss?

• Variable stress sweep settings (Min, Max, granularity).

• Run modes (explained in the following section).

The user has several controls to enable successful completion of Stressed Eye Cal including:

• Imodi ejwayelekile: This is the default and recommended run mode for Stressed Eye Cal. Preset and CMI/DMI stresses  are enabled, and BERT step response is captured and averaged, then passed to Seasim for analysis. This process is  repeated multiple times while sweeping stresses in the allowable range. Seasim computes eye width and eye height, and  TekRXTest software computes eye area then checks for convergence. A linear sweep of stresses (amplitude, SJ, DMI) is  performed to close the calibration.

• Imodi yokwenziwa ngesandla: Allows the user to manually enter SJ, DMI and amplitude values for Stressed Eye Cal. This mode is valuable  for “what-if” testing or debug to try discrete combinations of stresses for calibration closure.

• Exhaustive sweep mode: Allows the user to run Stressed Eye Cal for all combinations of amplitude, SJ and DMI within  the sweep range with defined step sizes. The user controls sweep ranges and step sizes as shown in Umfanekiso 14. The wider  the range and smaller the step size, the longer the run will take. This sweep ngeke ame when a converged eye is found,  rather it will perform a complete sweep of parameters and ranges defined by the user. The user may stop the sweep prior  to completion. If the user sees a combination of stresses that is desired then enable manual mode and perform stressed  eye verification with these settings to test for convergence.

imodi

Figure 14. User can control sweep ranges and step sizes.

16 | WWW.TEK.COM

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

TekRXTest software will iterate and adjust stresses until it finds a combination that results in an averaged eye width and eye height within the target range. In the example eboniswe ku Umfanekiso 15, a successfully converged stress calibration is shown in the green rows. In this example, averaged eye width = 2.929 ps and averaged eye height = 6.657 mV. These results were achieved using preset Q5, CTLE = 6 dB, amplitude = 726 mV, SJ = 1 ps, and DMI = 8.7 mV.

ukusetha ngaphambilini

Figure 15. Green rows show an example of a successfully converged stress calibration.

BERT Stress Calibration Best Practices

Before running BERT stress calibration, the user should ensure the following prerequisites are met to maximize accuracy and repeatability:

• Ensure scope and BERT have been powered on for ≥30 minutes. This ensures instruments are warmed up prior to  calibration.

• Ensure scope is running the latest versions of TekScope and waveform analysis software. To verify latest software  versions, please contact your local Tektronix field application engineer.

• On the scope with all RF cables disconnected, run signal path compensation (SPC) under UTLITITES ➞ INTRUMENT  CALIBRATION. SPC is recommended once every 30 days or if ambient temperature changes by >5 degrees C.

• Characterize PCIe 6.0 CEM fixture loss with VNA for TP2 ISI pair combination. For System Host, ISI trace pairs + RF cables  should measure ~24 dB loss @ 16 GHz (Add-in Card ~28 dB loss @ 16 GHz). Refer to PCIe 6.0 PHY Test Spec, Appendix C for more details.

WWW.TEK.COM | 17

PCI Express 6.0 CEM: BERT Stress Calibration Best Practices 

ISIFINYEZO SEZOBUCHWEPHESHE

The goal of BERT stress calibration is to find a combination of four adjustable stress parameters (ISI channel loss, SJ, DMI, amplitude) when combined with fixed stressors (i.e. RJ, CMI) that yields averaged eye height and eye width values within the specified range. These target ranges are:

• Eye width: 2.825 ps to 3.425 ps (nominally 3.125 ps).

• Eye height: 5.5 mV to 6.5 mV (nominally 6 mV).

The PCIe 6.0 PHY Test Specification specifies BERT stress calibration flow as follows:

1. Start with maximum ISI channel of 33 dB (HW ISI channel loss + embedded package model) for capturing the  step response.

2. Post process BERT step response with Seasim to find optimal preset and CTLE.

3. Post process BERT step response with Seasim to sweep adjustable parameters.

4. If a converged solution can be found, the calibration is successful. Record these values.

5. If no solution can be found with any combination of SJ, DMI, and amplitude, reduce ISI channel loss by one pair (nominally  0.5 dB step).

6. Repeat steps two through five until a solution converges within target eye width and eye height ranges.

VNA characterization of PCIe 6.0 fixtures along with RF cables is highly recommended prior to running BERT stress  calibration. TekRXTest software does provide a step-response based ISI channel verification, but this value may deviate  from a VNA-based measurement by several dB. The goal is to find a combination of ISI channel + RF cable loss that is ~24 dB  @ 16 GHz for System Host, ~28 dB @ 16 GHz for Add-in card. Once the ISI channel + RF cable combination is found, this is the  starting point for the TP2 portion of the calibration.

Ukuxazulula inkinga

In cases where TekRXTest does not successfully provide a converged BERT stress calibration at the end of TP2, below are  several areas to assess what changes should be made in the test setup or calibration settings:

• Check software revisions (scope, BERT, Sigtest, Seasim, etc.).

• Perform scope self-test and signal path compensation.

• Ensure scope channels are correctly deskewed <1 ps.

• Ensure coaxial connections in the signal path are tight and torqued to the proper value where appropriate.

• With no signal applied, measure pk-pk ACrms voltage on scope ATI Channel 1 and Channel 2 using auto measurements.  This ACrms value should be in the range of 1.1 to 1.3 mV. The measured ACrms mean value should also match between  Channel 1 and Channel 2 within a tight tolerance (≤.3 mV). If either condition is not met, ensure proper airflow into and out  of the scope to provide proper cooling. A minimum six-inch clearance is recommended on all sides of the scope. Ensure  instruments sitting in close proximity to the scope are not blowing hot air into the scope air inlet located on the bottom of  each scope unit.

Your Tektronix field application engineer is standing by to assist you. If you have followed this guide and are still not achieving  convergence during BERT stress calibration, reach out to your local Tektronix application engineer for assistance.

For more information on Tektronix’ PCI Express test solutions, please visit tek.com.

18 | WWW.TEK.COM

Ulwazi lokuxhumana:

Australia 1 800 709 465

I-Austria* 00800 2255 4835

AmaBalkan, Israel, South Africa kanye namanye amazwe we-ISE +41 52 675 3777

I-Belgium* 00800 2255 4835

Brazil +55 (11) 3530-8901

Canada 1 800 833 9200

Central East Europe / Baltics +41 52 675 3777

Central Europe / Greece +41 52 675 3777

I-Denmark +45 80 88 1401

I-Finland +41 52 675 3777

France* 00800 2255 4835

Germany* 00800 2255 4835

I-Hong Kong 400 820 5835

I-India 000 800 650 1835

I-Indonesia 007 803 601 5249

Italy 00800 2255 4835

Japan 81 (3) 6714 3086

E-Luxembourg +41 52 675 3777

I-Malaysia 1 800 22 55835

Mexico, Central/South America and Caribbean 52 (55) 88 69 35 25

Middle East, Asia, naseNyakatho Afrika +41 52 675 3777

Abase Netherlands* 00800 2255 4835

E-New Zealand 0800 800 238

eNorway 800 16098

IRiphabhulikhi Yabantu YaseShayina 400 820 5835

Philippines 1 800 1601 0077

Poland +41 52 675 3777

EPortugal 80 08 12370

IRiphabhulikhi yaseKorea +82 2 565 1455

Russia / CIS +7 (495) 6647564

Singapore 800 6011 473

Iningizimu Afrika +41 52 675 3777

i-Spain* 00800 2255 4835

I-Sweden* 00800 2255 4835

I-Switzerland* 00800 2255 4835

I-Taiwan 886 (2) 2656 6688

Thailand 1 800 011 931

United Kingdom / Ireland* 00800 2255 4835

USA 1 800 833 9200

Vietnam 12060128

* Inombolo yamahhala yase-Europe. Uma kungenjalo  

kufinyeleleka, shayela: +41 52 675 3777

I-Rev. 02.2022

Thola izinsizakusebenza ezibaluleke kakhulu ku- TEK.COM

Copyright © Tektronix. All rights reserved. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that  in all previously published material. Specification and price change privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix, Inc. All other trade names  referenced are the service marks, trademarks or registered trademarks of their respective companies.

082125 SBG 65W-74165-1